Vertical Automatic DA/DR Series Autoclave

FEATURES

  • Engineered safety: the lid and corners of the table top are covered with insulation to avoid burn injuries;
  • Exclusive patented locking device ‘with eight evenly distributed columns’ is safer and more productive;
  • Locking device: more reliable than ordinary locking devices;
  • Lid closing check: the machine cannot start if the lid does not close properly, by checking mechanically and electrically;
  • Real-time monitoring: the system will monitor the lid locking, pressure and temperature status in real time to ensure the safe and successful operation of the autoclave;
  • Cooling lock opening temperature: each programme can set a different lid opening temperature, especially for liquid programme, so that the user will not open the lid and get burnt when the products are still at high temperature;
  • Triple protection against lack of water: the bottom of the amber sterilisation chamber is equipped with three different drying zones. combustion protection devices: type of liquid expansion, type of copper temperature sensor and type of electrical I/O (water level sensor) to avoid misjudgement caused by one method;
  • Double overpressure protection;
  • Mechanical pressure gauge and electric sensor system, as soon as the pressure becomes abnormal, the system will cut off the power, release the pressure and sound an alarm;
  • Automatic Troubleshooting: the system will monitor the working condition and cut off the power, send alarm and report error information when any abnormal situation is detected;
  • Equipped with highly sensitive electrical protection system, effectively providing leakage, overcurrent and short-circuit protection.

 

Main technical specifications
ModelStandard GR60DA
With automatic drying function GR60DR
Capacity (L)60
Dimensions (L*W*H)(mm)DA:644x660x980
DR:644x831x980
Camera Size (Dia*H)(mm)φ390×505
Rated power (W)2900
SUS 304 Basket (Dia*H)(mm)(Note 1)(φ370×190) ×2
Power consumption220V+10% 16A 501IZ/60HZ
Temperature display accuracy0.1℃
Working environmentWorking environment 5℃~40℃ with relative humidity < 85%
Camera MaterialSUS304 stainless steel
Sterilization temperature range105℃~138℃
Sterilization time range1~6000 min (user-set, other mode can be set up to 480 min)
Heating temperature45℃~79℃
Warm-up time1~9999 min
Drying (DR only)Drying time: 1~300 min
Drying temperature: 80~160℃
Temperature range of the cooling lock opening40℃~99℃ for solids, waste, agar, self-separating40℃~80℃ (примечание 2) для жидких
Sterilization modeLiquid, liquid with a warming solid, wrapped tool, cloth, rubber, instant, waste-free, self-defining
Number of saved programsDA: 20 programs DR: 60 programs
Exhaust speed controlFully automatic internal unloading with a choice of 6 levels
The method of opening the lidReclines (up and down)
Water tankDA water tank: 2 water tanks for steam collection
DR: 1 water tank for steam collection + 1 water storage tank
Reception time range(Auto-start recording) 0~15 days
Pressure measuring range-0.1~0.5 MPa
Set pressure in the pressure valve0.29 Mpa
The controllerIntelligent microcomputer controller "Inspiration"
Safety device"Eight evenly distributed columns” locking device, electric double internal lock, lid closure check, dry burn protection, safety valve, water level sensor, overheating/pressure protection, temperature/pressure monitor, electrical protection (overcurrent/short circuit/leakage), cooling lock, lid against burns, automatic troubleshooting
Biosafety functionOption for a 0.2 microns thick bioprotective filter (basic model + S)
Note 2: The OPENING temperature of the cooling lock. refers to the local boiling point